tag:blogger.com,1999:blog-55221879729764781392024-03-15T17:02:28.874-07:00 Kulim Hi Tech Sdn Bhd Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.comBlogger131125tag:blogger.com,1999:blog-5522187972976478139.post-68113171021472896052023-10-26T00:22:00.000-07:002023-10-26T00:22:32.370-07:00Corrosion High Pressure High Temperature RCE Autoclave Cortest + Kulim Hi Tech Sdn Bhd <p></p><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEi_sO7vRdq066U1BlioGADHHDYxme8BB4sKRaMb9w864oKqmqpJPCyycYdMU1uFP9oTrx6laSscF6_-I37GwgpHfWAHTc6o9K9H8tQ1wgzp2KdWvOmaALQg3amEG6PifYrxcp8Vn5UIfDbgi1i8bsFHeEIJXiHduxZCToN_s4Y7c5oWFiM8TKWFADdQ7gSk/s4080/SGS%201.jpg" imageanchor="1" style="margin-left: 1em; margin-right: 1em;"><img border="0" data-original-height="4080" data-original-width="2296" height="400" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEi_sO7vRdq066U1BlioGADHHDYxme8BB4sKRaMb9w864oKqmqpJPCyycYdMU1uFP9oTrx6laSscF6_-I37GwgpHfWAHTc6o9K9H8tQ1wgzp2KdWvOmaALQg3amEG6PifYrxcp8Vn5UIfDbgi1i8bsFHeEIJXiHduxZCToN_s4Y7c5oWFiM8TKWFADdQ7gSk/w210-h400/SGS%201.jpg" width="210" /></a><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhcoLN9VrlTiz79-WnCFmb3aEPJ624wLAfpcrgVmyTdoyENDZ6OxvI6TheKExSQ8ti_6NJJpl3PNO7s6sIg-zR3MUdjSf_rSqhOmNgxLZkkYL_bNYsikxnHQfM7ejJxHo2VsDPIEvpEjX0ua3mjvOZjtRJc-XpGJHsHaPvz2oPshjnI8EB01DFgmx0Mc9cA/s4080/SGS%202.jpg" imageanchor="1" style="margin-left: 1em; margin-right: 1em;"><img border="0" data-original-height="4080" data-original-width="2296" height="400" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhcoLN9VrlTiz79-WnCFmb3aEPJ624wLAfpcrgVmyTdoyENDZ6OxvI6TheKExSQ8ti_6NJJpl3PNO7s6sIg-zR3MUdjSf_rSqhOmNgxLZkkYL_bNYsikxnHQfM7ejJxHo2VsDPIEvpEjX0ua3mjvOZjtRJc-XpGJHsHaPvz2oPshjnI8EB01DFgmx0Mc9cA/w225-h400/SGS%202.jpg" width="225" /></a></div><br /> <p></p><h3 style="background: rgb(255, 255, 255); border: 0px; box-sizing: border-box; color: white; font-family: ProximaNova-Regular, Helvetica, Arial, Lucida, sans-serif !important; font-size: 1.9vw; font-weight: 500; line-height: 1em; margin: 0px; outline: 0px; padding: 0px 0px 10px; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; color: black; font-size: 25.802px; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-size: 25.802px; font-weight: 700; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-family: Montserrat; font-size: 25.802px; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;">HPHT Autoclaves</span></span></span> </h3><p style="background: rgb(255, 255, 255); border: 0px; box-sizing: border-box; color: white; font-family: Montserrat, Helvetica, Arial, Lucida, sans-serif; font-size: 20px; margin: 0px; outline: 0px; padding: 0px 0px 1em; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; color: black; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-weight: 700; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;">Cortest autoclave systems are used for high pressure and high (and low) temperature (HPHT) test applications including: corrosion testing, electrochemical studies, corrosion inhibitor analysis, shear stress analysis, slow strain rate testing</span></span></p><p style="background: rgb(255, 255, 255); border: 0px; box-sizing: border-box; color: white; font-family: Montserrat, Helvetica, Arial, Lucida, sans-serif; font-size: 20px; margin: 0px; outline: 0px; padding: 0px 0px 1em; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; color: black; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-weight: 700; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;">Above image shown HPHT RCE autoclave.</span></span></p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-62264122855563669742023-08-21T23:43:00.000-07:002023-08-21T23:43:14.596-07:00Inhibitors Studies Bubble Tests- Gill 6,8,12 ACM Instruments - Kulim Kht<div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEi1Ob-1ZNPSlFUVBiDkV1CcMnUvl8XaXxsCvrd-yDid1mA4R04t5nTrLXff4pcnnhFOcJ34lMyf0oAX1cw7DPUxGOHTNO-KegK1wPMRsHu986n1lD5ex8IgjM1G5VFyDrq-sPxGjNZrca9QZzMF5VUY3BxhZMFC7bgW9aCJHvsII3SWRweayyBrDIzYAp1N/s4080/20230816_135250.jpg" imageanchor="1" style="margin-left: 1em; margin-right: 1em;"><img border="0" data-original-height="2296" data-original-width="4080" height="360" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEi1Ob-1ZNPSlFUVBiDkV1CcMnUvl8XaXxsCvrd-yDid1mA4R04t5nTrLXff4pcnnhFOcJ34lMyf0oAX1cw7DPUxGOHTNO-KegK1wPMRsHu986n1lD5ex8IgjM1G5VFyDrq-sPxGjNZrca9QZzMF5VUY3BxhZMFC7bgW9aCJHvsII3SWRweayyBrDIzYAp1N/w640-h360/20230816_135250.jpg" width="640" /></a></div><br /><p>Typical results are shown as metal has the highest corrosion rate but on
the addition of the inhibitor the rate drops to the lowest. The test should be
repeated at least three times and the results compared for scatter. The
percentage protection of an inhibitor is usually quoted this is defined as
100*(uninhibited rate-inhibited rate)/uninhibited rate. The bubble test software will
display this for all channels under test. The temperature of the test is important
as a whole set of percentage protection values is displayed with respect to
temperature.</p><p>Bubble Test facilitates running tests in multiples; tests are set up using a simple Bubble Wizard, a series of
LPR tests are collected over time giving Corrosion Rate results. Results can be printed at the end in many
ways including tables, lines and bar graphs. Cells (or Channels) are assigned together into Groups, each
group has an Add Inhibitor button, once pressed, end corrosion rates can be calculated at the end of the
test over 3 periods – prior to Add Inhibitor, after 2 hours and after 16 hours, this enables Percentage
Protection to be calculated.</p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-13824567994665590522023-08-21T23:36:00.000-07:002023-08-21T23:36:35.665-07:00Bubble Test - LPR - Inhibitors Gill 6,8,12 ACM Instruments Kulim Kht<p></p><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhjy0W7xnGJg_1VhjsYKFi_qIigJ3zl9AP6_6KKwrSIZKqKvzhqPf3o3HDs7P_f5JaGQKLirkC1TATPx2nbGG6A3yW64JOiDJp-sFAKYqD4Jza5OGYkNgmbBxH17MG-L4oiNR3AaK58L5SHeusVGDv5oIpRBCylY4pBktXQd_j57_6bFj9of8GB50ilhDuD/s4080/20230815_115159.jpg" imageanchor="1" style="margin-left: 1em; margin-right: 1em;"><img border="0" data-original-height="2296" data-original-width="4080" height="225" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhjy0W7xnGJg_1VhjsYKFi_qIigJ3zl9AP6_6KKwrSIZKqKvzhqPf3o3HDs7P_f5JaGQKLirkC1TATPx2nbGG6A3yW64JOiDJp-sFAKYqD4Jza5OGYkNgmbBxH17MG-L4oiNR3AaK58L5SHeusVGDv5oIpRBCylY4pBktXQd_j57_6bFj9of8GB50ilhDuD/w400-h225/20230815_115159.jpg" width="400" /></a></div><br />Testing of these chemicals is performed by makers and users. Makers set up
initial screening to not reject any potential good ones. Users set up initial
screening to reject any potential bad ones. Test metal should be the same metal
as that to be protected as should be the test fluid. The two most widely used
tests are LPR and Tafel plots. Both are ideally performed using a Gill 8 or Gill 12.
The use of a multiple channel instrument allows for replication to gain confidence
in the results. The Bubble Test software automatically groups the tests and prints
inhibition reports using the LPR method. After all the small LPR perturbations
have been made an optional Tafel sweep is performed to obtain the Tafel slopes.
AC impedance is less commonly used except with high resistance electrolytes
where the IR drop in the solution needs to be measured and overcome. <p></p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-18282469331153066422023-06-12T22:22:00.001-07:002023-06-12T22:25:17.379-07:00Potentiostat KHT ACM MultiPot 16.<iframe frameborder="0" height="270" src="https://youtube.com/embed/0cGwQdD2wCY" width="480"></iframe><div><br /></div><div><span style="background-color: white; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px;">The </span><a href="http://www.acminstruments.com/products/lab-instruments/pot-16" style="background: rgb(255, 255, 255); box-sizing: inherit; color: #da251d; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px; text-decoration-line: none; transition: all 0.2s ease 0s;" title="Click to view Pot 16 details">Pot 16</a><span style="background-color: white; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px;"> has 16 potentiostats built in, each at a fixed range i.e. not changed through the software. This allows very long tests of electrodes under potentiostatic control to be performed in complete certainty that the set potential will not be changed. Current and potential are recorded with time. Ideal for the study of cathodic disbondment. For more information click </span><a href="http://www.acminstruments.com/products/lab-instruments/pot-16" style="background: rgb(255, 255, 255); box-sizing: inherit; color: #da251d; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px; text-decoration-line: none; transition: all 0.2s ease 0s;" title="Click to view Pot 16 details">here</a><span style="background-color: white; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px;">.</span></div>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-88068098521202924012023-01-12T23:09:00.007-08:002023-01-12T23:14:34.736-08:00Surface Plasmon Resonance (SPR) Kulim Kht Bionavis Kinetics Titration<p></p><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEgVysWU5ouECojsnJB3UugMG-9V8BoU5Khz0tWWZsgPkCnlwfPu5DABpsiGU8gAYENzS57moZy-2Q_62SYDK2MCuLrnhSX3Yw4ODx5VMBM1qWeSMXH9bLyQH9CuBmtIg3pH2jAsMd1oB7w-RKyax0zxTxBAm83N2rGKoLIo14VButzzaEK7Gw1OVUMvFg/s1600/SPR%20410A.jpg" imageanchor="1" style="clear: left; float: left; margin-bottom: 1em; margin-right: 1em;"><img border="0" data-original-height="1516" data-original-width="1600" height="303" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEgVysWU5ouECojsnJB3UugMG-9V8BoU5Khz0tWWZsgPkCnlwfPu5DABpsiGU8gAYENzS57moZy-2Q_62SYDK2MCuLrnhSX3Yw4ODx5VMBM1qWeSMXH9bLyQH9CuBmtIg3pH2jAsMd1oB7w-RKyax0zxTxBAm83N2rGKoLIo14VButzzaEK7Gw1OVUMvFg/s320/SPR%20410A.jpg" width="320" /></a></div><span style="color: #ffa400; font-size: large;"><u>Introduction</u></span><p></p><p>
Surface Plasmon Resonance (SPR) is a popular real-time
label-free method to assess biomolecular interactions
especially in the fields of drug discovery and biosensor
development.
Multi-Parametric Surface Plasmon Resonance (MP-SPR)
scans a broad angular range and records whole SPR curve.
Hence from the same measurement, MP-SPR provides
more information about changes at the surface than
traditional SPR. Additionally, MP-SPR records also bulk
(solvent) effect simultaneously with the binding signal in
every channel. Thanks to the PureKinetics™ feature, bulk
effect can be separated from molecular binding events
in real-time, providing more reliable results and enabling
measurements, which are not possible using traditional
SPR instruments. PureKinetics™ enables measurement
of kinetics in altering dissociation buffers (Application
Note #147) To quantify affinity and kinetic parameters of the biomolecular interaction,
binding at not less than two concentrations of the analyte needs to be measured.
Concentrations need to be in the range close to the dissociation constant (KD),
and usually a few more concentrations are required, as affinity is typically
unknown. Conventionally, the dissociation rate is measured after each analyte
concentration. Additionally, if dissociation is slow, the remaining part of the
bound analyte is removed using a run of regeneration solution between
the sample injections. Ideal regeneration breaks the bonds between
the ligand and the analyte without deactivating the molecules on
the surface. However, such regeneration is sometimes difficult to achieve and also
the ligand is damaged.
The KineticTitration option allows the injection of samples sequentially and the
dissociation is measured only at the end of the whole run. This shortens the time
required for the whole measurement cycle (Figure 1).
KineticTitration is especially useful when:
• results are needed quickly
• there is less time available for assay optimization
(such as for finding a suitable regeneration solution)
• surface is difficult or impossible to regenerate without losing
the activity of the ligan<br /> </p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-59097278810549499032023-01-05T19:23:00.000-08:002023-01-05T19:23:40.022-08:00Weld Tester Kulim-KHT ACM Gill<iframe frameborder="0" height="270" src="https://youtube.com/embed/Pf4NmmOD5DY" width="480"></iframe><div><br /></div><div> ZRAs a weld tester is the
ideal instrument when testing segmented or mixed
metal systems. A full range of AC and DC
techniques are included for simultaneous testing of
all 1to16 ZRA. Each ZRA has the same frequency
response, count resistor ranges and power as a
standard Gill AC ZRA.</div>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-82509623283832302532022-12-21T22:58:00.001-08:002022-12-21T22:58:22.626-08:00Kulim Kht LPR Step for long Term Test ACM <div class="separator" style="clear: both; text-align: center;"><iframe allowfullscreen='allowfullscreen' webkitallowfullscreen='webkitallowfullscreen' mozallowfullscreen='mozallowfullscreen' width='320' height='266' src='https://www.blogger.com/video.g?token=AD6v5dyWvTsn3FMyS3bAwBFFAeiNnzXqMpmWg9NUids6BpGXhCsBe_6oaZM9C0d6rINyRm2Dwp-tyPKPhUO5ShNBRg' class='b-hbp-video b-uploaded' frameborder='0'></iframe></div><div class="separator" style="clear: both; text-align: center;"><br /></div><div class="separator" style="clear: both; text-align: center;"><span style="font-size: large;">This is a user friendly product and easy to use compare </span></div><div class="separator" style="clear: both; text-align: center;"><span style="font-size: large;">to others instruments. The software is complete package</span></div><div class="separator" style="clear: both; text-align: center;"><span style="font-size: large;"> and easy to collect data</span></div>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-36344046211547441842022-12-21T20:07:00.001-08:002022-12-21T20:07:55.284-08:00Merry Christmas and Happy New Year 2023<p></p><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhHQfGijkXM6DZJhhHJz1nNd7EWTZczfErjg2sAg6YyXstkLC_5J4hCqVKzr9M9Y1q-1BOXNHuqlefZUNCKfWISO4v0ssleATd73T9PniwoIO4XZXp0ofz7CI_RH-IBk20nACXmgWZ7bGbRYluCQLJf3nzQ_kY1pUu7amLW1pdQ9AGAv3AZCAAI5WnUUw/s562/Merry%20Christmas%202022.jpg" imageanchor="1" style="margin-left: 1em; margin-right: 1em;"><img border="0" data-original-height="562" data-original-width="447" height="640" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhHQfGijkXM6DZJhhHJz1nNd7EWTZczfErjg2sAg6YyXstkLC_5J4hCqVKzr9M9Y1q-1BOXNHuqlefZUNCKfWISO4v0ssleATd73T9PniwoIO4XZXp0ofz7CI_RH-IBk20nACXmgWZ7bGbRYluCQLJf3nzQ_kY1pUu7amLW1pdQ9AGAv3AZCAAI5WnUUw/w510-h640/Merry%20Christmas%202022.jpg" width="510" /></a></div><br /><span style="background-color: white; color: rgba(0, 0, 0, 0.9); font-family: -apple-system, system-ui, BlinkMacSystemFont, "Segoe UI", Roboto, "Helvetica Neue", "Fira Sans", Ubuntu, Oxygen, "Oxygen Sans", Cantarell, "Droid Sans", "Apple Color Emoji", "Segoe UI Emoji", "Segoe UI Emoji", "Segoe UI Symbol", "Lucida Grande", Helvetica, Arial, sans-serif; font-size: 14px;"> * Picture was painted originally with mouth by Konrad Gerber*</span> <p></p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-77562927596058347802022-12-20T23:13:00.000-08:002022-12-20T23:13:08.337-08:00Kulim Kht Weld Tester ACM Gill.<p> </p><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhUNySpGo_9sp47mO2P1Vmth-o4UMONpP8aRVlcdy7gdMuvUbX001LQJZrs3W1HQeJJGIV4Vew2WurmC-78M3gRh3QtYi-gVeBYoXGV-wjLUzoqapFsmlL9aoH8yLEi5o0eUrjmswXmu0l0abux07tWQPPc1bbtVdH1QQUp4L4lvH8QzsIbFo7TYgtRVQ/s4000/ACM%20Gill%20Weld%20Tester.jpg" imageanchor="1" style="margin-left: 1em; margin-right: 1em;"><img border="0" data-original-height="3000" data-original-width="4000" height="480" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhUNySpGo_9sp47mO2P1Vmth-o4UMONpP8aRVlcdy7gdMuvUbX001LQJZrs3W1HQeJJGIV4Vew2WurmC-78M3gRh3QtYi-gVeBYoXGV-wjLUzoqapFsmlL9aoH8yLEi5o0eUrjmswXmu0l0abux07tWQPPc1bbtVdH1QQUp4L4lvH8QzsIbFo7TYgtRVQ/w640-h480/ACM%20Gill%20Weld%20Tester.jpg" width="640" /></a></div><br /><p></p><p>Welds pose a particular problem. Each weld is a complex galvanic couple made
up of parent metal (e.g. a pipeline) a weld metal and an area of heat affected
metal. Under many circumstances the bulk of the anodic dissolution is within the
heat affected zone (haz). This very thin anode and large cathode of parent metal
is the worst possible case rapidly leading to failure. Conventional electrochemical
testing in which a sample of the weld is immersed and used as the working
electrode will only reveal a low overall net rate not the high rate in the heat
affected zone. The solution is to segment the sample weld cutting out test
electrodes of parent1 haz1 weld haz2 and parent2. These can then be coupled
together and the individual response from each section measured. The
instrument to do this is called a Weld Tester. This instrument is a Gill 8 or 12 built
to accommodate the extra four working electrodes per cell. Once the cell is
connected the full range of DC and AC tests are used to determine the
susceptibility of each component to corrosion.</p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-54148014182109312202022-11-28T22:32:00.000-08:002022-11-28T22:32:32.104-08:00Kulim Kht smartphone-based contact angle measurement instrument<p><span style="color: #660000;"><span style="font-size: large;"> </span><span style="font-family: Lora, serif;"><span style="font-size: large;">Conducting experiments to measure both static and dynamic</span></span></span></p><p><span style="font-family: Lora, serif;"><span style="color: #660000; font-size: large;"> (advancing and receding) contact angles</span></span></p><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjcrtUyrINy_SdeoUaMIfA_iYkctzjEUQY2z3sC41wnc_Cq3V18mEbnXD2to_QmDDGPz0ohY-8aDKyfoOaqvQtuzdj8bQpDiiEq1ovKhpTEA6OXQGuvl_K358eBkcI_uZx5_fj5OCpiMB6-LTsSh_Gud9UJ50YvijEhpGwvi_gq1b3et_KfADF6lZil5Q/s2048/KHT%20Droplet.jpg" imageanchor="1" style="margin-left: 1em; margin-right: 1em;"><img border="0" data-original-height="2048" data-original-width="2048" height="320" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjcrtUyrINy_SdeoUaMIfA_iYkctzjEUQY2z3sC41wnc_Cq3V18mEbnXD2to_QmDDGPz0ohY-8aDKyfoOaqvQtuzdj8bQpDiiEq1ovKhpTEA6OXQGuvl_K358eBkcI_uZx5_fj5OCpiMB6-LTsSh_Gud9UJ50YvijEhpGwvi_gq1b3et_KfADF6lZil5Q/s320/KHT%20Droplet.jpg" width="320" /></a></div><p><span style="font-family: Lora, serif; font-size: 20px;">In this study, a smartphone-based contact angle measurement instrument was developed. Compared with the traditional measurement instruments, this instrument has the advantage of simplicity, compact size, and portability. An automatic contact point detection algorithm was developed to allow the instrument to correctly detect the drop contact points. Two different contact angle calculation methods, Young-Laplace and polynomial fitting methods, were implemented in this instrument. The performance of this instrument was tested first with ideal synthetic drop profiles. It was shown that the accuracy of the new system with ideal synthetic drop profiles can reach 0.01% with both Young-Laplace and polynomial fitting methods. Conducting experiments to measure both static and dynamic (advancing and receding) contact angles with the developed instrument, we found that the smartphone-based instrument can provide accurate and practical measurement results as the traditional commercial instruments. The successful demonstration of use of a smartphone (mobile phone) to conduct contact angle measurement is a significant advancement in the field as it breaks the dominate mold of use of a computer and a bench bound setup for such systems since their appearance in 1980s</span></p><p><span style="font-family: Lora, serif; font-size: 20px;"></span></p><div class="separator" style="clear: both; text-align: center;"><br /></div><br /><p></p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-89370475217645980082022-11-09T23:13:00.001-08:002022-11-09T23:13:14.526-08:00Electrochemical Cell Set Kulim Kht ACM Bubble Test Set up<p></p><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/a/AVvXsEjC-bSXcKCQPz8__Z2KT7W1AueIjgFXCrFcFRq2GfRcHeX7X88ts0pkgvdUMSrabbviBrH5FsOdMS7E9QkLI3QIUKIlRtN5yIIdROU3nXL1Re0Z-r3PssLUHvuS_WZ_k3aHNBhl_96Au1dmDB7RnzBeR1XYSQNG8anhiF-CW6HV7_YPrQDJ7AWEt51Shg" style="margin-left: 1em; margin-right: 1em;"><img alt="" data-original-height="235" data-original-width="230" height="400" src="https://blogger.googleusercontent.com/img/a/AVvXsEjC-bSXcKCQPz8__Z2KT7W1AueIjgFXCrFcFRq2GfRcHeX7X88ts0pkgvdUMSrabbviBrH5FsOdMS7E9QkLI3QIUKIlRtN5yIIdROU3nXL1Re0Z-r3PssLUHvuS_WZ_k3aHNBhl_96Au1dmDB7RnzBeR1XYSQNG8anhiF-CW6HV7_YPrQDJ7AWEt51Shg=w392-h400" width="392" /></a></div><br /> <p></p><h3 style="background-color: white; box-sizing: inherit; font-family: Lato, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 1.5rem; margin: 0.75rem 0px 1.5rem; text-rendering: optimizelegibility;"><span style="color: red;">Electrochemical Cell Kit</span></h3><p><span style="background-color: white; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px;">The 1 Litre Quickfit type Electrochemical Cell is supplied with the following: Sealed 1 Litre Reaction Vessel; Platinum Band Type Auxiliary Electrode; Saturated Calomel Reference Electrode; Reservoir; Luggin Probe with Tubing; Platinum Mains Noise Reduction Electrode; Retort Stand; 0-105°C Thermometer; Glass Frit; Exhaust and a two Electrode Probe. The retractable probe fits in the centre of the cell and accommodates two standard 4.75cm</span><span style="background-color: white; box-sizing: inherit; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 12px; line-height: 0; position: relative; top: -0.5em; vertical-align: baseline;">2</span><span style="background-color: white; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px;"> screw on electrodes for galvanic C&V noise or 2 electrode LPR studies. The access hole in the centre of the cell which is normally used for the 2 electrode probe is supplied with an adapter for fitting of other probe types to suit your requirements. The adapter supplied with the Luggin probe can be removed in order to directly place the reference electrode into the cell.</span></p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-51763483087998382382022-11-09T20:29:00.000-08:002022-11-09T20:29:55.859-08:00Potentiostat Kulim Kht ACM Pot 16 ISO <p><br /></p><p></p><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhYKOVTXUJiFnL0viE6GA1LDh_ZtTx0Rlds13A-by9NB09MwMWSFwdJ5xjPbs6-FZMEAtPFnIxTQ3E6_OiAGy9wAFtDpM8yq-usaXr5ZDGO545_qbT784uBVJM6brg3VZh4-Ti2PAsNz2nRge4Q2hUsbftkHZWzepx4N9bIVh-aPlsXhwObselIe0WuCQ/s2592/DSC00013.JPG" imageanchor="1" style="clear: left; float: left; margin-bottom: 1em; margin-right: 1em;"><img border="0" data-original-height="1944" data-original-width="2592" height="480" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhYKOVTXUJiFnL0viE6GA1LDh_ZtTx0Rlds13A-by9NB09MwMWSFwdJ5xjPbs6-FZMEAtPFnIxTQ3E6_OiAGy9wAFtDpM8yq-usaXr5ZDGO545_qbT784uBVJM6brg3VZh4-Ti2PAsNz2nRge4Q2hUsbftkHZWzepx4N9bIVh-aPlsXhwObselIe0WuCQ/w640-h480/DSC00013.JPG" width="640" /></a></div><p></p><p><b><span style="color: red; font-size: medium;">DESCRIPTION <br /></span></b></p><p> Isolated Pot 16 has the same specifications as the Pot 16,
but each channel is isolated allowing samples to be in the
same electrolyte.
Fixed potentials for parallel potentiostatic tests.
This instrument provides 16 potentiostats for long term
testing of electrodes at a fixed potential.</p><p> Each channel has a fixed current measurement range,
selected on purchase for the type of testing to be
performed. The standard range is 350 mA with a
resolution of 5 micro Amps.
Output potential can either be fixed, selected by a two,
three or four position switch, or variable via a 10-turn
dialled knob. Extra optional potential selectors may be
added and channels grouped. </p><p>Typical configurations
would be all channels set by a two position switch
selecting 1.5V or 3V, or channels 1-8 set from -5V to +5V
via a 10 turn dialled knob and channels 9-16 set by their
own dialled knob. Most standard cathodic disbondment
tests (e.g. ISO15711, or ASTM G8, G42, G80, G95 or
CSA Z-245) require 1.05V, 1.5V, 3V or 3.5V; however a
galvanostat option can be added for the 3mA Australian
and New Zealand standards (AS-3862, AS/NZS4352). </p><p> Measurement of the current flowing from each potentiostat
and the voltage between RE and WE (as a check on
operation) is performed by a pair of 24 bit converters,
running at a maximum read rate of 4 channels per
second. </p><p> The software supplied allows for simple set up of data
filename and read rate and displays the latest data
recorded. The data is stored on the PCs hard disc as the
test progresses. In the event of a power cut the data will
be added to the earlier data on resumption of the power
supply.
An example of the use of a Pot 16 is in performing tests
on coated steel to study the likelihood of cathodic
disbondment at a fixed potential.
A dedicated and powerful instrument offering very cost
effective fixed potentiostats for long term measurement. </p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-40385652424379325882022-11-09T20:08:00.000-08:002022-11-09T20:08:21.348-08:00Kulim Kht ACM Weld Tester ZRA<p></p><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/a/AVvXsEiyblRAJx5fZcyeb4Ncns-76IIQ3dE-5q44c3ugN5XjUBYMWJPzgdIUNXtQm6YF0ummk63pEguaajuBf0qlrFNzDhzGu408A2nj77KOJ-qdsxeiHX3_3808pUO4YFj4HCBJ7rh-Nk-KnToNlq0sdC7qEhBqYg70wLOeT6JbbOSdjmVqDdXqYikm7KwLEg" style="clear: left; float: left; margin-bottom: 1em; margin-right: 1em;"><img alt="" data-original-height="164" data-original-width="300" height="175" src="https://blogger.googleusercontent.com/img/a/AVvXsEiyblRAJx5fZcyeb4Ncns-76IIQ3dE-5q44c3ugN5XjUBYMWJPzgdIUNXtQm6YF0ummk63pEguaajuBf0qlrFNzDhzGu408A2nj77KOJ-qdsxeiHX3_3808pUO4YFj4HCBJ7rh-Nk-KnToNlq0sdC7qEhBqYg70wLOeT6JbbOSdjmVqDdXqYikm7KwLEg" width="320" /></a></div><br /> <span style="color: red; font-size: x-large;">Weld Tester </span><p></p><p>With five independent ZRAs a weld tester is the
ideal instrument when testing segmented or mixed
metal systems. A full range of AC and DC
techniques are included for simultaneous testing of
all 5 ZRAs. Each ZRA has the same frequency
response, count resistor ranges and power as a
standard Gill AC ZRA.</p><p><span style="color: red; font-size: medium;"><br /></span></p><p><span style="color: red; font-size: medium;">Software Overview </span></p><p>At the heart of an ACM system is a Sequencer and Core Running application, now into Version 5
the emphasis is on reliability. Working in unison, Sequencer setups up a sequence of techniques
and Core Running collects data from a sequence of techniques. The Sequencer was designed to
be easy to use, with an intuitive interface, one that is common across the range from Data
Collection to Analysis; learning effort is kept to a minimum.</p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-32567584238323215792022-11-03T02:01:00.001-07:002022-11-03T02:01:18.166-07:00Kulim Kht CMC Surfactant analysis Easy <iframe width="480" height="270" src="https://youtube.com/embed/PEwzd4A-tYc" frameborder="0"></iframe>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-4895590504662883122022-11-01T23:30:00.001-07:002023-12-07T21:57:11.079-08:00Contact Angle meter, Modern Goniometer Kulim Kht<p>Optical Tensiometers are used in research, development and quality control
for the study of surfaces and interfaces. They will help you to characterize surfaces easily
and precisely, saving valuable time and money. Then Optical Tensiometers
support a wide range of applications from advanced research to quick quality control.
Thanks to the versatility, you can get the combination of features that best fits your
needs</p><p>Measurements
• Static contact angle
• Dynamic contact angle
• Roll-off angle
• Surface free energy (SFE)
• Surface tension
• Interfacial tension
• Batch contact angle
• Roughness corrected contact angle
• Interfacial rheology (viscoelasticity)</p><p><br /></p><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjUbrKFNvT0a_xU0rgt2Zp6knXgDYgO0gT-lEM60_C1HJ7J0H01GIp328aksXoEUxGlseRCOVxTZcmA0atyeArZR15Je2Wme-GqahDayYTBNKKwiKZ0TRKcRnTRJxNXCRkiXLN2RiYXy9c1WFP1-QTHuK4C0WBtzWmPfB6kkotJJHq9ch5x3LwEvC06QQ/s1024/Sirim%20Theta%20Flex.jpeg" style="clear: left; float: left; margin-bottom: 1em; margin-right: 1em;"><img border="0" data-original-height="768" data-original-width="1024" height="480" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjUbrKFNvT0a_xU0rgt2Zp6knXgDYgO0gT-lEM60_C1HJ7J0H01GIp328aksXoEUxGlseRCOVxTZcmA0atyeArZR15Je2Wme-GqahDayYTBNKKwiKZ0TRKcRnTRJxNXCRkiXLN2RiYXy9c1WFP1-QTHuK4C0WBtzWmPfB6kkotJJHq9ch5x3LwEvC06QQ/w640-h480/Sirim%20Theta%20Flex.jpeg" width="640" /></a></div><br /><p></p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-31552281021091293542022-08-16T01:45:00.001-07:002022-11-03T01:58:52.611-07:00Contact Angle Goniometer, Ball Calibration -Kulim Kht <iframe width="480" height="270" src="https://youtube.com/embed/dn8bUWkFCbI" frameborder="0"></iframe>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-59057972618757433992022-07-24T21:38:00.001-07:002022-07-24T21:38:59.179-07:00Kulim Kht ACM Cortest Rotating Cylinder Electrode (HPHT) Inhibitors Study<p> </p><h3 style="background: rgb(248, 248, 248); border: 0px; box-sizing: border-box; color: #333333; font-family: ProximaNova-Regular, Helvetica, Arial, Lucida, sans-serif !important; font-size: 1.9vw; font-weight: 500; line-height: 1em; margin: 0px; outline: 0px; padding: 0px 0px 10px; text-align: center; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-family: Montserrat; font-size: 25.954px; margin: 0px; outline: 0px; padding: 0px; text-decoration-line: underline; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-size: 25.954px; font-weight: 700; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEik7vWcpxnPYpt_FF55_6FqwQrow7y-m6h3UXyz-dkQbq6oJCcag1ZzSowMW2I8ON-6LQ6IfYYBnUVMIKUAFP2IeDPvm_l0bc9nRCQkTmq5wXcXMjDbgLgs3qBI_kpmYE952acml0WaW2RCb3PZQ71jRRrX-YxzQQahVuiOsednxs3C7v1prCVLc6o-xg/s4128/20190319_1711461.jpg" imageanchor="1" style="margin-left: 1em; margin-right: 1em;"><img border="0" data-original-height="3096" data-original-width="4128" height="240" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEik7vWcpxnPYpt_FF55_6FqwQrow7y-m6h3UXyz-dkQbq6oJCcag1ZzSowMW2I8ON-6LQ6IfYYBnUVMIKUAFP2IeDPvm_l0bc9nRCQkTmq5wXcXMjDbgLgs3qBI_kpmYE952acml0WaW2RCb3PZQ71jRRrX-YxzQQahVuiOsednxs3C7v1prCVLc6o-xg/s320/20190319_1711461.jpg" width="320" /></a></div><br /><span style="background: transparent; border: 0px; box-sizing: border-box; color: black; font-size: 25.954px; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><br /></span></span></span></h3><div class="separator" style="clear: both; text-align: center;"><div class="separator" style="margin-left: 1em; margin-right: 1em; text-align: center;"><img alt="" data-original-height="810" data-original-width="1080" height="240" src="https://blogger.googleusercontent.com/img/a/AVvXsEjjfPKV-NAqAEzpS9PUqhkkz-M7wrRrhtgG3fOXcujag4PwOjWpnA2oZo_ATxQE9__yNHEn81IEr19dpWQIgbkSpFUb0MsAAtv1vvGfbzHWm0n7TiZexL6PoI-884WOYX8nxQzTtcOe3ojR9jkAwY3lWJaHaOjy3lwZOKXNcJQRZmJt62TiijbAuFabUA=w302-h240" width="302" /></div></div><h3 style="background: rgb(248, 248, 248); border: 0px; box-sizing: border-box; color: #333333; font-family: ProximaNova-Regular, Helvetica, Arial, Lucida, sans-serif !important; font-size: 1.9vw; font-weight: 500; line-height: 1em; margin: 0px; outline: 0px; padding: 0px 0px 10px; text-align: center; text-size-adjust: 100%; vertical-align: baseline;"></h3><h3 style="background: rgb(248, 248, 248); border: 0px; box-sizing: border-box; color: #333333; font-family: ProximaNova-Regular, Helvetica, Arial, Lucida, sans-serif !important; font-size: 1.9vw; font-weight: 500; line-height: 1em; margin: 0px; outline: 0px; padding: 0px 0px 10px; text-align: center; text-size-adjust: 100%; vertical-align: baseline;"><span style="background-color: transparent; color: black; font-family: Montserrat; font-size: 25.954px; font-weight: 700; text-decoration-line: underline;">Cortest Rotating Cylinder</span></h3><h3 style="background: rgb(248, 248, 248); border: 0px; box-sizing: border-box; color: #333333; font-family: ProximaNova-Regular, Helvetica, Arial, Lucida, sans-serif !important; font-size: 1.9vw; font-weight: 500; line-height: 1em; margin: 0px; outline: 0px; padding: 0px 0px 10px; text-align: center; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-family: Montserrat; font-size: 25.954px; margin: 0px; outline: 0px; padding: 0px; text-decoration-line: underline; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-size: 25.954px; font-weight: 700; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; color: black; font-size: 25.954px; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"> Electrode & ACM Gill AC</span></span></span></h3><h3 style="background: rgb(248, 248, 248); border: 0px; box-sizing: border-box; color: #333333; font-family: ProximaNova-Regular, Helvetica, Arial, Lucida, sans-serif !important; font-size: 1.9vw; font-weight: 500; line-height: 1em; margin: 0px; outline: 0px; padding: 0px 0px 10px; text-align: center; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-family: Montserrat; font-size: 25.954px; margin: 0px; outline: 0px; padding: 0px; text-decoration-line: underline; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-size: 25.954px; font-weight: 700; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; color: black; font-size: 25.954px; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"> for Data Measurements</span></span></span></h3><div><span style="background: transparent; border: 0px; box-sizing: border-box; font-family: Montserrat; font-size: 25.954px; margin: 0px; outline: 0px; padding: 0px; text-decoration-line: underline; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-size: 25.954px; font-weight: 700; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; color: black; font-size: 25.954px; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><br /></span></span></span></div><p style="background: rgb(248, 248, 248); border: 0px; box-sizing: border-box; color: #2f2f2f; font-family: "Open Sans", Arial, sans-serif; font-size: 14px; margin: 0px; outline: 0px; padding: 0px; text-align: center; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; color: black; font-family: Montserrat; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;">The Rotating Cylinder Electrode (RCE), as described in ASTM G185, is a cylindrical shaped specimen attached to a stirring shaft which rotates at a specific speed. The specimen is electrically isolated from the other components in the system and the corrosion on its outer surface is measured. At various speeds, the surface of this electrode specimen is subjected to various boundary layers and flow effects. At Cortest, we have developed a system that can accommodate any size specimen. This system can be fitted to any existing autoclave and used at temperatures up to 220°C (428°F) and pressures of 15.1 MPa (2,200 PSI). Ports can be included in the lid for electrochemical probes.</span></p><p style="background: rgb(248, 248, 248); border: 0px; box-sizing: border-box; color: #2f2f2f; font-family: "Open Sans", Arial, sans-serif; font-size: 14px; margin: 0px; outline: 0px; padding: 0px; text-align: center; text-size-adjust: 100%; vertical-align: baseline;"><br /></p><p style="background: rgb(248, 248, 248); border: 0px; box-sizing: border-box; color: #2f2f2f; font-family: "Open Sans", Arial, sans-serif; font-size: 14px; margin: 0px; outline: 0px; padding: 0px; text-align: center; text-size-adjust: 100%; vertical-align: baseline;">Coupon holder options Rotating</p><p style="background: rgb(248, 248, 248); border: 0px; box-sizing: border-box; color: #2f2f2f; font-family: "Open Sans", Arial, sans-serif; font-size: 14px; margin: 0px; outline: 0px; padding: 0px; text-align: center; text-size-adjust: 100%; vertical-align: baseline;"> Cage</p><p style="background: rgb(248, 248, 248); border: 0px; box-sizing: border-box; color: #2f2f2f; font-family: "Open Sans", Arial, sans-serif; font-size: 14px; margin: 0px; outline: 0px; padding: 0px; text-align: center; text-size-adjust: 100%; vertical-align: baseline;"> Rotating Coupon</p><p style="background: rgb(248, 248, 248); border: 0px; box-sizing: border-box; color: #2f2f2f; font-family: "Open Sans", Arial, sans-serif; font-size: 14px; margin: 0px; outline: 0px; padding: 0px; text-align: center; text-size-adjust: 100%; vertical-align: baseline;"> Rotating Cylinder Electrode </p><p style="background: rgb(248, 248, 248); border: 0px; box-sizing: border-box; color: #2f2f2f; font-family: "Open Sans", Arial, sans-serif; font-size: 14px; margin: 0px; outline: 0px; padding: 0px; text-align: center; text-size-adjust: 100%; vertical-align: baseline;">3 and 4 Point Bend Fixtures
Static Coupon<br /><br /></p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-13559677369952539062022-07-24T20:55:00.000-07:002022-07-24T20:55:22.525-07:00High Pressure High Temperature Rotating Cylinder Electrode (RCE) Kulim Kht Cortest Incp.<p> </p><p><br /></p><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjPDqmA5zxa5NTMN9eT3290V_J7od0a6NVYr4OuAF8To6UfO03-E3mBJnMQnLi1uypQUzrRy1Li-96r-CnlBByHhsPVmaInv8AXFQFqGHyw6APwN3w9AdqjuI4mUH5vznrHH5hFrWQsCyqDxXmXEVVQgCUUDe7s6yeckR9bo5mHgCL-_vPKB_lKB0D8BA/s4128/20190319_1710502.jpg" imageanchor="1" style="margin-left: 1em; margin-right: 1em;"><img border="0" data-original-height="3096" data-original-width="4128" height="300" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjPDqmA5zxa5NTMN9eT3290V_J7od0a6NVYr4OuAF8To6UfO03-E3mBJnMQnLi1uypQUzrRy1Li-96r-CnlBByHhsPVmaInv8AXFQFqGHyw6APwN3w9AdqjuI4mUH5vznrHH5hFrWQsCyqDxXmXEVVQgCUUDe7s6yeckR9bo5mHgCL-_vPKB_lKB0D8BA/w400-h300/20190319_1710502.jpg" width="400" /></a></div><br /><p><br /></p><h3 style="background: rgb(255, 255, 255); border: 0px; box-sizing: border-box; color: white; font-family: ProximaNova-Regular, Helvetica, Arial, Lucida, sans-serif !important; font-size: 1.9vw; font-weight: 500; line-height: 1em; margin: 0px; outline: 0px; padding: 0px 0px 10px; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; color: black; font-size: 25.954px; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-size: 25.954px; font-weight: 700; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-family: Montserrat; font-size: 25.954px; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;">RCE Autoclaves</span></span></span> </h3><p style="background: rgb(255, 255, 255); border: 0px; box-sizing: border-box; color: white; font-family: Montserrat, Helvetica, Arial, Lucida, sans-serif; font-size: 20px; margin: 0px; outline: 0px; padding: 0px 0px 1em; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; color: black; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; font-weight: 700; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;">Cortest autoclave systems are used for high pressure and high temperature (HPHT) test applications including: corrosion testing, electrochemical studies, corrosion inhibitor analysis, shear stress analysis, slow strain rate testing, and more.</span> When paired with one of our <span style="background: transparent; border: 0px; box-sizing: border-box; margin: 0px; outline: 0px; padding: 0px; text-decoration-line: underline; text-size-adjust: 100%; vertical-align: baseline;"><a href="http://cortest.com/project/inline-and-belt-driven-stirrers/" style="background: transparent; border: 0px; box-sizing: border-box; color: black; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;">Magnetic Stirrers</a></span> or <span style="background: transparent; border: 0px; box-sizing: border-box; margin: 0px; outline: 0px; padding: 0px; text-decoration-line: underline; text-size-adjust: 100%; vertical-align: baseline;"><a href="http://cortest.com/project/hpht-circulating-flow-loops/" style="background: transparent; border: 0px; box-sizing: border-box; color: black; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;">Circulating Flow Loop Systems</a></span>, our Autoclaves may be used for dynamic testing applications.</span></p><p style="background: rgb(255, 255, 255); border: 0px; box-sizing: border-box; color: white; font-family: Montserrat, Helvetica, Arial, Lucida, sans-serif; font-size: 20px; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;"><span style="background: transparent; border: 0px; box-sizing: border-box; color: black; margin: 0px; outline: 0px; padding: 0px; text-size-adjust: 100%; vertical-align: baseline;">Cortest has been building safe and reliable autoclave systems for over 40 years and offers standard system designs as well as custom configurations to meet the most demanding of applications. Each pressure vessel is designed in accordance to the ASME Section VIII Div. 1 standard. PED, CSA, CE, DOSH</span></p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-43618186035400236542022-04-21T23:24:00.000-07:002022-04-21T23:24:46.020-07:00Kulim Kht ACM Pot 24 with 8 variable potentials, each controlling a group of 3 channels<p> </p><p></p><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjYvboKHW1XPUeEgJqR7kaXemOEQAkcLKdIifa8s0_JCBsOAlchqYPkgj29s09RaTEzulJfjHum_nK9DxkUZHIQnJzSLw4pZI5bkcqDOmISOphXX_HyH3A-BKKRnsQSUjZIdCBO1TzOHp8jlmGEOjRY6qH8CYmU9nZCpeoEkkgVvXMUQ9HTB09Hcj9nkQ/s3840/IMG-20220301-WA0000.jpeg" imageanchor="1" style="clear: left; float: left; margin-bottom: 1em; margin-right: 1em;"><img border="0" data-original-height="2160" data-original-width="3840" height="147" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjYvboKHW1XPUeEgJqR7kaXemOEQAkcLKdIifa8s0_JCBsOAlchqYPkgj29s09RaTEzulJfjHum_nK9DxkUZHIQnJzSLw4pZI5bkcqDOmISOphXX_HyH3A-BKKRnsQSUjZIdCBO1TzOHp8jlmGEOjRY6qH8CYmU9nZCpeoEkkgVvXMUQ9HTB09Hcj9nkQ/w229-h147/IMG-20220301-WA0000.jpeg" width="229" /></a></div><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjmVDvw_m4ShMLFb88kO5cj3AAsbxeYaA0pzjfqEume_O2MgZ5Ut0BxPrU6cHuBoa2ppTQ6Y606qxfdrHMaEuzDPULwAqfjtNq47BnERvS4a1L2iePr4JVF33neWq3sOnywUdTI83proxBqpqnAPOj64V4v2QRykJ97CY2QYdNZpxds47Sz3fOLpszXjA/s3840/IMG-20220301-WA0004.jpeg" imageanchor="1" style="clear: left; float: left; margin-bottom: 1em; margin-right: 1em;"><img border="0" data-original-height="2160" data-original-width="3840" height="129" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjmVDvw_m4ShMLFb88kO5cj3AAsbxeYaA0pzjfqEume_O2MgZ5Ut0BxPrU6cHuBoa2ppTQ6Y606qxfdrHMaEuzDPULwAqfjtNq47BnERvS4a1L2iePr4JVF33neWq3sOnywUdTI83proxBqpqnAPOj64V4v2QRykJ97CY2QYdNZpxds47Sz3fOLpszXjA/w228-h129/IMG-20220301-WA0004.jpeg" width="228" /></a></div><p></p><p><strong style="background-color: white; box-sizing: inherit; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px;">Pot 24</strong><span style="background-color: white; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px;"> with 8 variable potentials, each controlling a group of 3 channels</span></p><p style="background-color: white; box-sizing: inherit; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px; margin: 1rem 0px;">Long term testing at fixed potentials is vital in many branches of corrosion, coatings, cathodic protection and electrochemistry. The Pot 16 offers 16 individual potentiostats (number of channels can be determined by your needs). Each channel has a fixed current measurement range, selected on purchase for the type of testing to be performed. The standard range is 350mA with a resolution of 5µA.</p><p style="background-color: white; box-sizing: inherit; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px; margin: 1rem 0px;">Output potential can either be fixed, selected by a switch, or variable via a 10-turn dialled knob. Extra optional potential selectors may be added and channels grouped. Typical configurations would be all channels set by a two position switch selecting 1.5V or 3.5V, or channels 1-8 set from -5V to +5V via a 10 turn dialled knob and channels 9-16 set by their own dialled knob. Most standard Cathodic Disbondment tests (e.g. ISO15711, or ASTM G8, G42, G80, G95 or CSA Z-245) require 1.05V, 1.5V, 3V or 3.5V; however a galvanostat option can be added for the 3mA Australian and New Zealand standards (AS-3862, AS/NZS4352). Please specify which standards you require when quoting. We will ensure your Pot 16 complies.</p><p style="background-color: white; box-sizing: inherit; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px; margin: 1rem 0px;">A single dialled knob is fitted as standard, to allow a wide range of potentials to be selected for all channels. Additional dialled knobs may be specified to allow different potentials on groups of channels. In some cases a variable potential is not desirable, in which case, switches may be specified to allow selection between fixed potentials e.g. -1.5V & -3.5V.</p><p style="background-color: white; box-sizing: inherit; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px; margin: 1rem 0px;">Alternatively up to 4 fixed potentials may be ordered and allocated between the 16 channels at no extra cost. For example channels 1 to 8 could be set to -1.5V, channels 9 to 11 set to -3V, channel 12 to 14 set to -3.5V and 15 to 16 set at -4V. These potentials are set on internal 25 turn trimming potentiometers.</p><p style="background-color: white; box-sizing: inherit; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px; margin: 1rem 0px;">To increase the value of these tests both current and voltage are monitored and stored using very high accuracy internal voltmeters. The voltage should stay constant at the set potential but if this is not the case it is usually an indication of problems with the cell (for example a bubble in the Luggin probe).</p><p style="background-color: white; box-sizing: inherit; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px; margin: 1rem 0px;">The Pot 16 can easily be customised to the customers specifications. Higher current output, more channels, less channels, waterproof enclosure, alternative connectors and battery powered option are all possible. The Pot 16 is infinitely variable. It is designed to accommodate the needs and specifications of the end user.</p><h4 style="background-color: white; box-sizing: inherit; color: #424753; font-family: Lato, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 1.25rem; margin: 0.75rem 0px 1.5rem; text-rendering: optimizelegibility;">Standards:</h4><ul style="background-color: white; box-sizing: inherit; color: #424753; font-family: Raleway, Helvetica, Tahoma, Geneva, Arial, sans-serif; font-size: 16px; margin-bottom: 1.5rem; margin-left: 1.5rem; margin-top: 1.5rem; padding: 0px;"><li style="box-sizing: inherit;">ASTM G8, G42, G80, G95 and their variants</li><li style="box-sizing: inherit;">AWWA C-210, API RP 5L7</li><li style="box-sizing: inherit;">CSA Z-245.20-02 and Z245.21-02</li><li style="box-sizing: inherit;">BS EN10289 : 2000(E)</li><li style="box-sizing: inherit;">ISO15711 Method A & B</li><li style="box-sizing: inherit;">ISO-21809-2 and ISO-21809-3</li><li style="box-sizing: inherit;">NACE RP0394</li><li style="box-sizing: inherit;">Shell DEP 31.40.30.31-Gen and Shell DEP 31.40.30.32-Gen</li><li style="box-sizing: inherit;">DIN 30 671 (1987)</li><li style="box-sizing: inherit;">NF A49-711, NF A49-716</li><li style="box-sizing: inherit;">09-SAMSS-089 ARAMCO</li><li style="box-sizing: inherit;">GBE/CW6:Part 1 and BGC-PS-CW6:Part 1</li><li style="box-sizing: inherit;">ARCO China</li><li style="box-sizing: inherit;">AS-3862, AS/NZS4352:1995 Method A & B. <strong style="box-sizing: inherit;">(Pot 16 with Galvanostat option or <a href="http://www.acminstruments.com/products/lab-instruments/gal-16" style="background: transparent; box-sizing: inherit; color: #da251d; text-decoration-line: none; transition: all 0.2s ease 0s;" title="Click to view Gal 16 details">Gal 16</a> Only)</strong></li></ul>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-31822155444446026942021-10-04T22:51:00.002-07:002022-11-03T02:00:24.440-07:00Meniscus Contact Angle Kulim Kht in Bahasa Malaysia<iframe width="480" height="270" src="https://youtube.com/embed/IrkborD-F3E" frameborder="0"></iframe>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-90295843615851241342021-10-04T00:30:00.003-07:002022-11-03T02:00:46.343-07:00Kulim KHT Data Receiver Connection <iframe style="background-image:url(https://i.ytimg.com/vi/SV2bQ6_W8n0/hqdefault.jpg)" width="480" height="360" src="https://youtube.com/embed/SV2bQ6_W8n0" frameborder="0"></iframe>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-90565400705649592242020-10-06T00:56:00.001-07:002023-12-07T22:04:00.729-08:00Ring Du Noüy ring Measurement Kulim Kht<p style="text-align: center;"> </p><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEisrLVUZCECdrqTIHcm5Jw9JTPagFWsBl6lwFfNdyAgOADEX1hByegLmDrulpw6aWJavYgpR3c5X13SKced4nEGfYhKYTDDDM9VPRQO8qkelOwiJbD_uGQ5aq7DQGOkdm-F51gu37wztLR5R3WQZ5aAlQOt2Z1Wez-D_ZqTnE7bj_Ra_uekbqcfIQQe1jJM/s1688/Du-%20Nouy%20Spec..jpg" imageanchor="1" style="margin-left: 1em; margin-right: 1em;"><img border="0" data-original-height="1446" data-original-width="1688" height="274" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEisrLVUZCECdrqTIHcm5Jw9JTPagFWsBl6lwFfNdyAgOADEX1hByegLmDrulpw6aWJavYgpR3c5X13SKced4nEGfYhKYTDDDM9VPRQO8qkelOwiJbD_uGQ5aq7DQGOkdm-F51gu37wztLR5R3WQZ5aAlQOt2Z1Wez-D_ZqTnE7bj_Ra_uekbqcfIQQe1jJM/s320/Du-%20Nouy%20Spec..jpg" width="320" /></a></div><br /><p></p><p>Optional<span style="font-size: medium;"><b> platinum/iridium ring</b> </span>is supplied in a protective transport case. The ring should be cleaned
regularly after use and treated carefully to avoid deformation. Dimensions: R (ring) = 9.58 mm, r
(wire)= 0.185mm.
A good way to store the ring is to immerse it in ethanol by first hanging it on a hook attached to a
lid, which is then attached to a beaker filled half-full with ethanol. Then before using the ring again
flush it with pure ethanol and water, or burn it with a Bunsen burner (ca. 1000 oC flame) for a short
time (glowing yellow-red for a few seconds). In this way the ring will keep its form better and will
not deform as easily as by putting it back to the wooden box after each measurement.
The ring that is used must have R/r ratios of around 52 or 26. This is because when the Du Noüy
ring is used, a correction factor has to be applied. The software includes correction factor for rings
with R/r of 52 or 26. You can read more about the correction in the Theory section of this manual. </p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-80286930779697559632020-10-05T23:20:00.001-07:002023-12-07T22:06:49.652-08:00Contact Angle Tilting stage Kulim Kht <p></p><div class="separator" style="clear: both; text-align: center;"><br /></div><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjD_zqeBGVHFcT_uxO6z4J2ULW2RrR_1v_KfwTYpYZ2N4aAc79Su7PkL5CfTqOoEQOLwAT_W38wt3ZXpDDzrgs6wmM-cvESaFY21zQl8EjqpbRC3u9_-HNmy21CEriyN5w2fNwvSkc2H4HoAWaGLNoYR_3l3D2rEKvXYHWJVh1Sw4vWzcLIlb-sKjOeQAp2/s1834/Droplet-Kulim%20Kht.png" imageanchor="1" style="margin-left: 1em; margin-right: 1em;"><img border="0" data-original-height="1834" data-original-width="1828" height="320" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjD_zqeBGVHFcT_uxO6z4J2ULW2RrR_1v_KfwTYpYZ2N4aAc79Su7PkL5CfTqOoEQOLwAT_W38wt3ZXpDDzrgs6wmM-cvESaFY21zQl8EjqpbRC3u9_-HNmy21CEriyN5w2fNwvSkc2H4HoAWaGLNoYR_3l3D2rEKvXYHWJVh1Sw4vWzcLIlb-sKjOeQAp2/s320/Droplet-Kulim%20Kht.png" width="319" /></a></div><br /><div class="separator" style="clear: both; text-align: center;"><br /></div><br /> <b><span style="font-size: medium;">Manual tilting stage
</span></b>Manual tilting stage enables the study of dynamic contact angle i.e. advancing and receding contact
angles. Manual tilting stage is attached to similarly as standard sample stage. <p></p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-724230677212679492020-10-05T23:01:00.002-07:002023-12-07T22:08:42.571-08:00Solar Contact Angle Measurement Kulim Kht <p></p><div class="separator" style="clear: both; text-align: center;"><br /></div><div class="separator" style="clear: both; text-align: center;"><br /></div><div class="separator" style="clear: both; text-align: center;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjzFV0FhOl8_ADzbul_i5y8QhI1GH0ERS6OQklM-RQjOjapcA4zhdLL1pvM03DB2CemdPWwXAbpW7Do_zx2ZOOQffCuKCpL4jSbzWv98PgDWpZXp6Ayvky55RKC5JSUWBFqwA0ocUe2VoH8lqs5ZZ89ZuQtTeTsAFsfV2zXZLsF33znpAbrCWPChnhWLkzc/s675/Kht%20measurement_1-1024x576-700x350-min.jpg" imageanchor="1" style="margin-left: 1em; margin-right: 1em;"><img border="0" data-original-height="456" data-original-width="675" height="216" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEjzFV0FhOl8_ADzbul_i5y8QhI1GH0ERS6OQklM-RQjOjapcA4zhdLL1pvM03DB2CemdPWwXAbpW7Do_zx2ZOOQffCuKCpL4jSbzWv98PgDWpZXp6Ayvky55RKC5JSUWBFqwA0ocUe2VoH8lqs5ZZ89ZuQtTeTsAFsfV2zXZLsF33znpAbrCWPChnhWLkzc/s320/Kht%20measurement_1-1024x576-700x350-min.jpg" width="320" /></a></div><br /><div class="separator" style="clear: both; text-align: center;"><br /></div><br /><span style="font-family: arial;"> <span style="font-size: medium;"><b>Solar cell research and development</b></span> </span>is an active field and the
efficiency of the solar cells is constantly increasing. Contact angle
measurements can be utilized in many stages of solar cell research
and solar panel manufacturing as well as in quality control of
solar cells. Anti-reflective coatings are used in all solar panels
to improve efficiency and lifetime of the solar cells. Study of the
surface properties of these coatings is commonly done by using
contact angle measurements. Ready solar panels are often too
large to be fitted into a laboratory contact angle instrument. is a great tool for quality control and testing of solar
panels due to its portability and unlimited sample size.<p></p>Kulim Hi-Tech Sdn Bhdhttp://www.blogger.com/profile/12428445220167857152noreply@blogger.com0tag:blogger.com,1999:blog-5522187972976478139.post-17671416826124247062019-10-02T23:26:00.002-07:002022-11-03T02:05:38.723-07:00KULIM Kht Advantages of optical tensiometry Contact Angle Measurements.<br />
<h4>
<span style="background-color: #f8f8fb; clear: left; display: inline; font-family: "roboto" , "arial" , sans-serif; margin-bottom: 1em; margin-right: 1em;"><a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhG1Lf_vPdXu4LjGCoNDnblWClsJXj0TfrNwva4z65_XPr670CqL0OMR-XEceBvdvbXa-PNrq8vUUbFbvqnTL9zdjcRdM5QS0zMlz92vsVdYo3roG-eZ_ghETtO_WHGWm7IwX50E5jqwNeB/s1600/20191003_124026.jpg" imageanchor="1" style="clear: left; display: inline !important; margin-bottom: 1em; margin-right: 1em;"><span style="color: #0b5394;">Advantages of optical tensiometry</span></a></span><span style="color: magenta;">DEMO ON RUBBER GLOVE MOULD.</span></h4>
<h4>
<span style="background-color: #f8f8fb; color: #6c6e70; font-family: roboto, arial, sans-serif; font-size: xx-small;"><a href="https://www.biolinscientific.com/service-and-support?id=5311468711&group=attension&item=5384752463" style="border-bottom: 1px solid rgba(128, 130, 133, 0.25); box-sizing: inherit; color: #00a1e4; text-decoration-line: none; transition: color 0.2s ease-in-out 0s, background 0s ease 0s, opacity 0s ease 0s, border 0s ease 0s;">Optical tensiometry</a> is the best choice when liquid-solid interactions and/or solid properties (such as<span style="box-sizing: inherit;"> </span><a href="https://www.biolinscientific.com/measurements/surface-free-energy" style="border-bottom: 1px solid rgba(128, 130, 133, 0.25); box-sizing: inherit; color: #00a1e4; text-decoration-line: none; transition: color 0.2s ease-in-out 0s, background 0s ease 0s, opacity 0s ease 0s, border 0s ease 0s;" title="Surface Free Energy">surface free energy</a>)</span><span style="background-color: #f8f8fb; color: #6c6e70; font-family: roboto, arial, sans-serif; font-size: xx-small;"> are</span></h4>
<div style="background-color: #f8f8fb; box-sizing: inherit; color: #6c6e70; font-family: Roboto, Arial, sans-serif; line-height: 1.77778; margin-bottom: 24px; padding-bottom: 0px; padding-top: 0px;">
<h3>
<span style="font-size: xx-small;"> measured on samples that are not homogeneous and regular in size and shape.</span><span style="font-size: xx-small;"> The main benefits are following:</span><span style="font-size: xx-small;"> Small liquid volume demanded (only some microliters)</span><span style="font-size: xx-small;"> Solid substrate can be asymmetrical (e.g. rubber glove mould)</span><span style="font-size: xx-small;"> Both sides of the sample don’t need to be similar (coating and surface modification characterization)</span><span style="font-size: xx-small;"> Contact angle mapping over the surface enables</span><span style="box-sizing: inherit; font-size: xx-small;"> </span><span style="font-size: xx-small;">heterogeneity and</span><span style="box-sizing: inherit; font-size: xx-small;"> </span><span style="font-size: xx-small;">cleanliness characterization</span><span style="font-size: xx-small;"> Roughness correction possible e.g. by using 3D Topography module</span></h3>
</div>
<div class="separator" style="clear: both; text-align: center;">
<a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEgqv45M4MatFPIy1Y_o-gPb-kQEcugfMKndN3dGrhsvv27QJyWGT847dput2zYALUWr0VmMSAcVT5ukmglkSYm2b6ed4EvyDCHE-ah1vHiSiVf7nKuFOLes8JCzWnNsgu51NXq69KCxuzdM/s1600/20191003_124143.jpg" imageanchor="1" style="clear: left; float: left; margin-bottom: 1em; margin-right: 1em;"><img border="0" data-original-height="1247" data-original-width="1600" height="496" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEgqv45M4MatFPIy1Y_o-gPb-kQEcugfMKndN3dGrhsvv27QJyWGT847dput2zYALUWr0VmMSAcVT5ukmglkSYm2b6ed4EvyDCHE-ah1vHiSiVf7nKuFOLes8JCzWnNsgu51NXq69KCxuzdM/s640/20191003_124143.jpg" width="640" /></a></div>
<div class="separator" style="clear: both; text-align: center;">
<a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEiHPU6S5IVjcY4ft-gz8YHPEfjdIxfz2docUbJRlqQo_QbRpvg4tHyZQlJrTkLZUlO_5ZixFCRzFtWy5tQRM6rVkZWkcAIAtpk5QWgnde7-ymLctDPUI9EFDLPRYUHaz219F9EMjTM-_o74/s1600/20191003_123917.jpg" imageanchor="1" style="clear: left; float: left; margin-bottom: 1em; margin-right: 1em;"><img border="0" data-original-height="1188" data-original-width="1600" height="474" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEiHPU6S5IVjcY4ft-gz8YHPEfjdIxfz2docUbJRlqQo_QbRpvg4tHyZQlJrTkLZUlO_5ZixFCRzFtWy5tQRM6rVkZWkcAIAtpk5QWgnde7-ymLctDPUI9EFDLPRYUHaz219F9EMjTM-_o74/s640/20191003_123917.jpg" width="640" /></a></div>
<div class="separator" style="clear: both; text-align: center;">
<a href="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEgsfy1FALryIHNCjWA1Ro_YbbaMpGtEZwDnFUdH4YBypDoXUEvx3OejOP3DxyEdvfvfdKvdHkOqYOMuzBvffgQYGv7SGLThog6K7vmXKwtiDg_2vV1PbdPuyLES_mewD76mP_exJhtNIyUh/s1600/20191003_124026.jpg" imageanchor="1" style="clear: left; float: left; margin-bottom: 1em; margin-right: 1em;"><img border="0" data-original-height="1329" data-original-width="1600" height="528" src="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEgsfy1FALryIHNCjWA1Ro_YbbaMpGtEZwDnFUdH4YBypDoXUEvx3OejOP3DxyEdvfvfdKvdHkOqYOMuzBvffgQYGv7SGLThog6K7vmXKwtiDg_2vV1PbdPuyLES_mewD76mP_exJhtNIyUh/s640/20191003_124026.jpg" width="640" /></a></div>